How to improve the data accuracy of electronic universal testing machine
Release time:
2024-09-09
Whether the sample is clamped correctly during the test of the electronic universal testing machine, incorrect clamping will affect the accuracy of the test results.
In addition to paying attention to the daily maintenance of the electronic universal testing machine, we should also regularly check other aspects of the universal testing machine. If we encounter a large deviation in the test data, we should first check and eliminate the following situations to improve the data accuracy of the electronic universal testing machine:
1. The test speed of the electronic universal testing machine is not standard during the test. Too fast or too slow will lead to different test results.
2. Is the sample clamped correctly during the test? Incorrect clamping will affect the accuracy of the test results.
3. Is there a zeroing operation before the electronic universal testing machine? If the test is started without zeroing, the test results will definitely be inaccurate.
4. Is the force sensor damaged by external impact or over-capacity use? The damage of the sensor will affect the accuracy of the data.
5. After the sensor has been used for a certain period of time, the strain undergoes elastic deformation and does not return. In fact, this strain deformation is very small. It is recommended to recalibrate the coefficient of the force sensor once a year and correct the data.
Latest News
Recommend a flue pressure testing machine
The precise automatic control and data acquisition system realizes the full digital adjustment of data acquisition and control process. It conducts tests and provides test data according to GB, ISO, JIS, ASTM, DIN, BS and other international standards. It automatically completes the test parameters such as the maximum bearing capacity and strength of the material.
2024-09-09
Spring tension and compression testing machine software function introduction
After the test, the data can be saved automatically or manually. After the test, the maximum force, upper and lower yield strength, Hysteresis loop method, stepwise approximation method, non-proportional extension strength, tensile strength, compressive strength, elongation strength at any point, Elongation at any point with fixed load, elastic modulus, elongation, maximum value, minimum value, average value, net energy of the peeling interval, Reentry energy, total energy, bending modulus, breakpoint displacement x% load, breakpoint load X% displacement, etc.
2024-09-09
High and low temperature spring tension and compression fatigue testing machine features
Embedded high-performance single-chip CPU controller; dot matrix LCD; full Chinese operation display; multi-key operation; automatic memory of parameter settings, automatic connection test at startup, front pickup
2024-09-09
Recommend a locomotive coil spring fatigue testing machine
Main test waveforms: sine wave, triangle wave, square wave, trapezoidal wave, ramp wave and program control wave, etc.
2024-09-09
Introduction of New Electro-hydraulic Pulsation Fatigue Testing Machine
The electro-hydraulic pulsating fatigue testing machine adopts a unique crank-connecting rod rocker mechanism to achieve pulsating sinusoidal wave loading on the test piece. It adopts the best energy-saving design and its energy consumption is one twentieth of that of traditional fatigue equipment!
2024-09-09
Precautions for adjusting the zero point of hydraulic universal testing machine
When the cylinder is not raised, the test force displays a negative value, which is normal. Please do not adjust the zero point.
2024-09-09